4Fusion®

The 4Fusion® shape memory quadripodal implant is designed
to give compressive stabilization in 4 - corner fusion
arthrodesis.

ADVANTAGES

Stability of assembly:

  • Efficient and constant compression of fragments
  • Instant rigid fixation
  • Notched legs for improved anchoring
  • Allows for immediate motion

Low Profile Implant:

  • Bone conserving
  • No impingement with radius in flexion
  • Open design allows for bone graft placement

Principle of Use/Performance

  • Store in freezer (below 32°F) before use
  • The implant recovers its shape at body temperature
  • After removal from the freezer, the general work time is approximately 1.5 to 2 minutes
  • Supplied in a sterile vacuum package

Brief Technique

  • After the proper excision of the scaphoid, denude the joint spaces of cartilage.
  • Perform a reduction and pin with a K–wire.
  • Utilizing the reamer guide and reamer, prepare the fusion surface.
  • Fill any fusion gaps with cancellous bone graft
  • Use the reamer guide and prepare the holes for the implant legs.
  • Remove the implant from the freezer (work time is 1.5 to 2 minutes).
  • Utilizing the implant support and impactor, insert the implant
  • The implant will compress at body temperature.

Shape Recovery

Remove retainer prior to impaction

MULTI-DIRECTIONAL COMPRESSION: Crown compresses and legs compress

Implant SIZING

Cat.# Thickness Reamer Inter-Axis
A
Inter-Axis
B
Leg Length
QUA-4F-14 1.7 x1.5m 2.0mm 12mm 13mm 14mm
QUA-4F-16 1.7 x1.5m 2.0mm 12mm 13mm 16mm
QUA-4F-18 1.7 x1.5m 2.0mm 12mm 13mm 18mm